Operational Specifications | ||
Description | Nominal | Range |
Power | 125 Watts | 75-150 Watts |
Current | 12 amps (DC) | 7-14 amps(DC) |
Operating Voltage | 11 Volts (DC) | 9.5-12.5 Volts(DC) |
Ignition Voltage | 17 Kilovolts(system dependant) | |
Temperture | 150℃ (Maximum) | |
Life Time | 1000 hours(500 hour warranty) |
Initial Output at Nominal Power | |
F= UV Filtered Output/UV=Enhanced Output | |
Description | PE125BF |
Peak Intensity | 300x10³ Candelas |
Radiant Output* | 17 Watts |
UV Output* | 0.8 Watts |
IR Output* | 10 Watts |
Visible Output* | 1500 Lumens |
Color Temperature | 5600° Kelvin |
Peak Instabilities | 4% |
Beam Geometry | 4.5°/5°/6° |
* These values indicate total output in all directions. Wavelengths = UV<390 nm, IR>770 nm,
Visible: 390 nm-770 nm
* Beam Geometry Defined as half angle at 10% PTS after 01/100/1000hours
Description | Visible Output | Total Output* |
6 mm aperture | 1050 Lumens | 9.5 Watts |
8 mm aperture | 620 Lumens | 5.6 Watts |
1. Lamp must not be operated with window facing upwards within 45° of vertical.
2. Seal temperature must not exceed 150°.
3. Current/power regulated power supplies and Excelitas lamp housing units are recommended.
4. Lamp must be operated within recommended current and power range. Over powering may lead to arc instability, hard starting and premature aging.
5. Hot mirror assembly is available for IR filtering.
6. Cermax® Xenon lamps are much safer lamps to use than their quartz xenon arc lamp equivalents. However, caution must be practiced when operating lamps because they are under high pressure, require high voltage, reach temperatures up to 200℃ , and their IR and UV radiation can cause skin burns and eye damage. Please read the Hazard Sheet included with each lamp shipment